<< Test and Measurement Products <<

RFB-2000 Radiation Force Balance  NEW!

Designed specifically for easy Ultrasound Power measurements on medical devices, the RFB-2000 requires little maintenance and no complex correction factors, and is compliant with regulatory requirements including AIUM-NEMA UD-2, UD-3, IEC 61161, and with associated IEC standards. Automation assures reliable means to validate Acoustic Output measurements as recommended by AIUM-NEMA document UD-2. Data Sheet [810 kB]



Features

  • Transparent, top-loading configuration
  • No membrane or coupling gel required
  • Unaffected by water evaporation
  • Can be user-calibrated to a standard masses
  • Automatic temperature correction
  • Accommodates a variety of targets
  • Selectable averaging to optimize measurement time and precision
  • Automatic and manual measurement modes
  • Automated logging of configuration, cal and measurement data
  • Three means for automatic control of user equipment
  • Includes calibration certificate based on acoustic power
  • USB interface, optically isolated, Win XP compatible

Performance Specifications

  • Power Ranges (selectable): 1 mW* to 2W with standard absorbing target, up to 20W (RFB-CTK) and up to 100W** (RFB-BTK)
  • Display Resolution: >4 digits
  • Typical measurement uncertainty*: <5% at 95% confidence level
  • Measurement Cycle Time: default 4s (15 measurements per minute)
  • Noise (Std. Dev. for N=1): <3 mW
  • Maximum beam diameter: 5 cm
  • Low Frequency Limit: 1 MHz (usable at lower frequency with decreased accuracy)

          *  Extended averaging may be required, depending on local vibration and power level

          ** Depending on beam parameters

Optional Measurement Kits

  • RFB-BTK Brush target for HIFU - Up to 100W
  • RFB-CTK Cone target kit for Physiotherapy - Up to 20W
  • RFB-VLK Vacuum lid for in-tank degassing

Accessories

  • Water-compatible vacuum pump
  • Set of standard Weights





Weight (Tank & Stand) - 7.5Kg plus 6.5Kg water maximum

This specification is subject to change without notice

<< Test and Measurement Products <<

592 Weddell Drive, Suite 7, Sunnyvale, CA 94089, Tel: (408) 745-0383, Fax: (408) 745-0956
Copyright © 1995-2002 Onda Corporation. All Rights Reserved
.