Page 11 - HydrophoneHandbook_eReader_Ed2_20151208
P. 11

Overview

An equivalent electrical circuit representing the hydrophone
measurement is shown below:

+

Q Rx Cx Cc Ca Ra V                                                      V

                                                                    -              t

    Where:                                                             Oscilloscope
    Q = Charge produced from applied pressure
    Rx = Resistance of piezo
    Cx = Capacitance of piezo
    Cc = Capacitance of cable
    Ca = Capacitance of amplifier circuit
    Ra = Resistance of amplifier circuit
    V = Measured voltage

Reference:
http://web.stanford.edu/class/me220/data/lectures/lect10/lect_6.html

                                                                                            9

                                                                       0NDA Corporation